Brief description:The MPC-9604 is an advanced, multi-detector alpha/beta counter system designed for applications requiring high sample volumes and low detection limits.

  • Manufacturer strength

    Manufacturer Strength
  • Valid warranty

    Valid Warranty
  • Quality assurance

    Quality Assurance

Products

PRODUCT CATEGORY

Detailed introduction

MPC-9604It is an advanced, multi-detectoralpha/betaCounter system designed for applications requiring high sample volumes and low detection limits.


              





  • Manual, four-detector sample counter
  • space defense probe
  • Field-replaceable detector windows
  • 4-inch thick low-background lead shield
  • 2-inch standard sample pan diameter
  • low background counter
  • Radon daughter compensation for air filters
  • GasPr safety gas protection and gas loss monitoring
  • Four 2.25-inch round pancake-type flow proportional detectors
  • Independent sample measurements without mutual interference
  • Modular design, expandable to 48 channels
  • information +

    • Download link MDS brochure
    • Download link MDS Brochure (A4)
  • More information +


    The MPC-9604 is designed with performance features designed to simplify setup, operation and maintenance.

    •ofThe connection is AC power,USBcomputer control andP-10gas supplyYes, it consists ofGas pro™Monitor and stop counting when a gas failure is detected.

    •straightGuan’s control software provides simple data access and reporting capabilities, allowing for limiteddrastically reducetraining and setup time. Each detector has its own sample drawer for independent sample loading and control, true“counting toMDA”, andwith“ MDAcount”toOptimize throughput.

    •Maintenance does not require rear access to the detector and electronics modules, which can be accessed from the front panel. The detector window can easily be10pointscan be replaced within the clock without interrupting the operation of other detectors。

  • Specifications +


    Efficiency top (about 1mm from the window);

    Others: bottom of sample tray
    source top 1/8"  1/4" 5/16"
    Pu-239 35% 31% 25% 24%
    Sr/Y-90 55% 47% 40% 39%
    Tc-99 35% 33% 28% 26%


    Typical efficiency (distributed/point source)
    source efficiency source efficiency
    Po-210 31% Am-241 31%
    Sr/Y-90 47% Th-230 31%
    Tc-99 33% Cs-137 32%
        

    Background
    Configuration
    Guaranteed value (cpm)
    Typical value (cpm)
    Alpha Beta Alpha Beta
    Shield and defend <0.1 <1.0 <0.05 <0.7


    Instrument specifications
    Category Specifications
    Sample size 2inch diameter (5.1cm)
    Detector Uniformity Effective area exceeds 95%
    Capacity 4samples
    Instrument crosstalk

    α Yes β <0.1%
    β Yes α <0.1%

    Depth 1/8,1/4or5/16inches
    Data export VISTA 2000 Software
    interface computer
    Counting mode onlyα Or firstαAgainβ orα/βAt the same time
    Ping

    α: Ping Xian<1.5%/100V; Tsubocho>1000V
    β: Ping Xian<2.5%/100V; Pingchang>200V

    power supply 110-240 V AC, 50-60 Hz 
    <2 A @ 110 V AC, <1.0 A @ 240 V AC
    Gas flow and pressure

    Usually set to60cc/min,With a critical flow orifice plate;
    Usually the outlet pressure level of the pressure reducing valve10psi.

    window Aluminized, 80 µg/cm2
    Operating temperature range 10ºC-40ºC(50ºFto104ºF)
    Humidity 20 - 90%No condensation
    shield 4inches (10.16cm)Low background lead shielding
    Specifications and dimensions

    16inches highx 32inches widex 22inches deep
    (47cmhigh x 95.3cmwide x 55.9cmdeep)

    weight 1200pounds (544 kg), Ship 1400pounds(635 kg)


    Factory measurement conditions
    • Determination of 4π efficiency using NIST traceable area source @ 1/8"
    • Measure background at natural background 8 µR/hr
  • Ordering information +


    Customer must provideAC power, P-10 gas supply, calibration source, and a suitable surface on which to place each MPC-9604. The supporting surface must be able to safely support the 1,200-pound weight of each MPC-9604.

    Model Description
    ORT-MPC 9604
    includePICConnecting cables, two-stage regulator and tubing, and 4 inserts of the same size (specify1/8inches,1/4inches or5/16inches)
    Options
    ORT-MDS-CTRL
    MDScontrol package. includeVista 2000software,RS232/485interface, null modem cable, gas regulator
    ORT-MDS-RACK
    Heavy-duty rack assembly holds twoMPC-9604unit
    ORT-VISTA 2000
    software package
    PIC-700-0007 1/8inch sample pan(per set100a)
    PIC-700-0008 1/4inchesSample tray(everyset100a)
    PIC-700-0009 5/16inchesSample tray(everyset100a)
    PIC-120-0008 Ultra thin(80μg/cm2)detector window
    PIC-120-0011 mobile cart

Product consultation