A count rate monitoring system is included in this module. This system can be used for external switching when modulating the pulse count rate. When the count rate is equal to or higher than the count rate selected by the monitor count rate discriminator, a voltage >+2 V will appear on the signal monitor connector. When the count rate is lower than the selected value, 0 V will appear on the connector. Threshold is selected by front panel switch.
The type 9302 amplifier is partially AC coupled, with input and output impedances of 50Ω respectively. The amplifier's rise time is typically 3 ns when the gain is set to 200, and <2 ns when the gain is set to 20.
The discriminator portion of the Model 9302 is a leading edge discriminator whose level is adjustable from 50 mV to 1 V.