ORTEC FoodGuard 2is a quantitative solution for the determination of food radioactivity via high-resolution gamma spectroscopy with a high-purity germanium detector. It is designed to simplify and simplify sample handling in busy laboratory environments. theFoodGuard-2-AKAdd-on kits are easy to install and add-on to anyORTECGamma spectroscopy systems have been usedGammaVisionsoftware. without affectingGammaVisionWith the system running, it can be installed in minutes and, once calibrated, provides users with standardFoodGuard-2full functionality of the system. This greatly simplifies operation.
The resulting workstation is a ready-to-use, fully functional system that enables sophisticated analysis of food, water and other agricultural products that contain radioactive materials that may produce any form of gamma ray emission following an accident or incident.
FoodGuard-2Workstation software included from international regulatory organizations(such as the U.S. Food and Drug Administration and the World Health Organization)acceptable radionuclide activities(Such as137Cs、134Csand131I)An example of a restriction table. These tables can also be customized with additional nuclides and different limits to meet your specific needs.
in order to makeFoodGuard-2With the system running, supervisors can calibrate, set sample types, alarm limits and report outputs in just a few simple steps. For regular“High sample throughput”To use, the operator enters some key sample information using an interactive user interface, places the sample in one of the included marlin cups, and begins counting.
During the count, the software provides feedback such as energy spectra and alarm notifications, then presents a professionally formatted report that can be shared or archived. The purpose of this operation is for high sample throughput and clear results.
FoodGuard-2-AKWhat's in the extras?
FoodGuard-2-AKAdd-ons include6a2liter of marin sample cup and a check source, including a2Liter full(40K)Marin Cup. This beaker is suitable for end cap diameters less than or equal to83mmHPGedetector. The scale needs to be copied using(2Marin Cup)Standard, if not available, can be usedOptional angle efficiencyConversion software from other similar geometries(Such as1Marin Cupor bottle)Transfer scale.
In addition to sample containers and inspection sources,FoodGuard-2-AKAdd-on package includes two software projects:
- FG-2-B32 FoodGuard-2 Software version1.0.11or higher.
- 797230 Connection upgrade kit version7.02.03or higher.
System minimum detection concentration
MDCThe implementation depends on the specifics of the installation system:Lead shielding specifications,HPGeRelative efficiency of detector, size of sample/Quantity.
In the table below,MDCDesignated for use in a typical system20%detector and2Liter of liquid in a marlin glass.(Not guaranteed on any specific system). For larger detectors and longer counting times,MDCswill be lower(better)。
Scales for additional sample densities and geometries are availableGammaVisionCreate. Once created, operators can easily load different scale calibrations for different sample types.
| measured 137Cs MDC* | counting time |
| 0.75 Bq/L | 10 minutes |
| 0.5 Bq/L | 30minutes |
| *UseISO 11929 method calculation. | |