Product model:

Brief description:High performance, multi-function MCA/counter/timer/count rate meter in single-wide NIM 200MHz counter performance

  • Manufacturer strength

    Manufacturer Strength
  • Valid warranty

    Valid Warranty
  • Quality assurance

    Quality Assurance

Products

PRODUCT CATEGORY

Detailed introduction

928 with Multichannel Buffer MCB, Analog to Digital Converter ADC, Counter, Timer, Ratemeter
  • High performance, multi-function MCA/counter/timer/count rate meter in single-width NIM
  • 200 MHz counter performance, Log/Lin alarm count rate meter, fast (1.25 μs) 16k MCA, USB2.0 connection
  • Intuitive front-panel emulator software and optional programmer support toolkit
  • *Flexible internal interconnection
  • Programmable external inputs and outputs
  • Network capable of passing ORTEC CONNECTIONS architecture
  • Easy to integrate into measurement systems and experiments

The ORTEC EASY-NIM 928 instrument suite provides new methods of nuclear pulse counting and pulse height analysis for a variety of nuclear science applications, providing customers with cost-effective flexibility, reliability and ease of use. Each product in the series comes with integrated hardware in the form of a compact, single-width NIM module; software in the form of a "front panel emulator" for Windows environments, and an optional toolkit that enables users to integrate the hardware into their own experimental or instrumentation systems.
  • information +

    • Download link 928 Brochure
    • Download link 928 Brochure (A4)
    • Download link 928 manual
  • More information +


    Hardware introduction
    Hardware features of the EASY-NIM 928 series include:

    • multi-channel analyzer
    • 200 MHz pulse counter
    • Timer with internal or external time base function
    • Count rate meter with adjustable alarm threshold
    • 6 configurable universal inputs
    • 2 configurable outputs (rear panel)

    Software introduction
    A separate "front panel emulator" provides display and user control functions for the MCA and counter/timer/count rate meter sections. Through the design architecture of ORTEC Connections, the EASY-NIM 928 hardware suite can be controlled from these applications installed locally or remotely over the network. An optional programmer's toolkit allows custom, user-written programs and controls.

    List of EASY-NIM models
    Model 928 is a combination of quad counter/dual timer and multi-channel analyzer functionality. Model 928-MCB is just a MCA, while models 928-COUNT2 and 928-COUNT4 are just counters/timers. With the exception of the 928-MCB, each model is equipped with two programmable outputs and six inputs for connection internally or to external devices that may be assigned to various signal sources.

    All in one

    independent

    928

    MCB/Four-way counter/
    Combination of dual timer modules

    928-MCB

    onlyMCB

    928-COUNT2

    only2counters/1timers

    928-COUNT4

    only4counters/2timers


  • Specifications +


    multi-channel buffer (928and928-MCBmodel)

    ADC

    Successive approximation type with sliding scale linearization

    resolution

    16,384channels, software selectable16,384、8,192、4,096、2048、1024and512。

    Dead time for each event

    2 μsIncludes memory transfer.

    integral nonlinearity

    before the dynamic range99%,<+0.025%。

    differential nonlinearity

    before the dynamic range99%,<±1%。

    gain instability

    <+50 ppm/°C。

    Dead time correction

    The software can be based onGedcke-Halemethod orZDTLive time correction for extended live time
    Interval correction, which monitors the count rate and adjusts for dead time during count rate fluctuations.

    data memory

    512 kb。


    input and output

    ADCinput

    Accepts positive unipolar, positive gated integrator, or0to+10 VBipolar analog pulses that are predominantly positive within the dynamic range;+12 V;0.25Arrive30 μsThe semi-Gaussian time constant of3Arrive30 μsthe time constant of the gated integrator shape, or the delay line shape, where the width>0.25 μs。Zin =1kΩ DC coupling. No internal delays,BNCconnector.

    ADCdoor

    OptionalTTLEnter. The computer can choose conform mode, anti-conform mode, or off. The signal must occur before and exceed the pulse peak0.5-μs;BNCconnector.
    Z
    in =1kΩ。

    OptionalTTLinput

    The computer can choose conform mode, anti-conform mode, or off. The signal must occur before and exceed the pulse peak0.5-μs;BNCconnector.Zin =1kΩ。

    PUR

    Stack rejection input; acceptTTLSignal; the signal must occur before peak detection.Zin =1kΩ。BNCconnector.

    busy

    Busy input used by the live time correction circuit. acceptTTLSignal; the signal must occur before peak detection.Zin =1kΩ。BNCconnector.


    Default
    Presets can be set to automatically stopMCBcollection. Additionally, counters and timers can be configured to work withMCBSynchronous start
    and stop.

    real time/living time

    20Multiple of milliseconds.

    area of interest

    Peak count/Points counting.

    Data overflow

    When any channel exceeds231-1Data collection will be terminated.

    peak uncertainty

    Acquisition stops when the statistical or counting uncertainty of the user-selected net peak reaches a specified value.

    nuclideMDA

    When the user specifiesMDAThe minimum detectable activity (MDA) value reaches the specified value, data collection will stop.


    counter/timer/Count rate meter (928、928-COUNT2、928-COUNT4)

    Count rate

    200 MHz。

    Number of counters

    4(2one in928-COUNT2middle).

    Number of timers

    2(1one in928-COUNT2middle).

    Counting capacity

    32Bit.

    Pulse pair resolution

    7 ns。


    input and output

    Counter input1、2、3、4

    Fast analog signal inputs are available on the front panelBNCThe connector accepts amplitudes up to ±5 Vof analog or digital pulses. Pulses are counted when they exceed the discriminator threshold. The computer selects to trigger on positive or negative slope. The threshold can be determined by computer at-1.6 Vto+3 VAdjust within the range, the step size is1.5 mV(minimum pulse30 mV). Computer optional50Ω or1000Ω input impedance, DC coupled. At the discriminator threshold, the minimum input pulse width is3.5 ns. The count rate is200 MHz。

    Input1、2、3、4、5、6

    front panelBNCInput acceptsTTLinput signal. Each input is<0.8 Vis considered low when the>2 Vis considered high level. The input impedance is1000Ω is grounded. Each input can be configured as described previously for counter and timer settings.

    Output1、2

    rear panelBNCoutput, used as a generalTTLoutput terminal.<0.4 Vis low level.>+2.4 Vis high level. The output can be driven50 Ω. As mentioned before, the outputs can be configured via software.


    counter/Timer software control

    Reading control

    Real-time values of timers and counters can be read from the computer at any time. Additionally, each counter or timer has a128LiteralFifo(First-out) buffer used to hold the previous final counter result when a counter or timer is reset. UseFifo, the counter can be configured to reset after the timer reaches a preset value. The current counter value is stored in itsFifo, so the computer can read consecutive values without worrying about losing results because the computer is slow.


    Electrical and Mechanical (all)

    USB-2.0

    StandardUSBConnection is via the included10ft. of cable implemented.

    indicator

    ADCinstructionsADC-1activity.
    Counter: Will light up if any of the four counters are counting.

    MCBSoftware control

    (Use the includedMAESTROperform operations - See datasheet for details. )
    ADC LLD
    : The computer can be accessed from0Arrive100% of full scale for control.
    ADC ULD
    : The computer can be accessed from0Arrive100% of full scale for control.
    Live Time: Correction Gedcke-Hale,ZDT。
    Door:Conform, anti-conform, close.

    Required power supply

    +6 V, 350 mA; -6 V, 225 mA; +12 V, 100 mA; –12 V, 100 mA。

    weight

    Net weight 0.9Kilogram (2pounds).Shipping weight 2.25Kilogram (5pounds).

    Size

    NIMStandard, single width,3.43 x 22.13cm (1.35 x 8.714inches) front panel, conforming toDOE/ER0457Trequirements.


  • Ordering information +


    Model

    Description

    928

    MCB/Four-way counter/Combination of dual timer modules. includeMAESTRO MCASimulation software, front panel software andUSBcable.

    928-MCB

    Instrumental onlyMCBsection (no counter/timer function). includeMAESTRO MCAsimulation software andUSBcable.

    928-COUNT2

    module only2counters/1timer section (noneMCBfunction). Includes front panel software andUSBcable.

    928-COUNT4

    module only4counters/2timer section (noneMCBfunction). Includes front panel software andUSBcable.