Detailed introduction

ORTEC 671 Spectroscopy Amplifier
  • High-performance energy spectroscopy for all types of radiation detectors (Ge, Si, scintillation; proportional counters)
  • Compact single-width NIM module
  • Optional triangle and Gaussian filters double the time constant of resolution
  • Automatic noise discriminator on stack rejector and baseline restorer eliminates screwdriver adjustments
  • Automatic baseline recovery rate for performance at low and high count rates
  • Differential inputs reduce ground loop noise

The ORTEC Model 671 High Performance Spectral Amplifier is ideally suited for use with germanium, Si (Li) and silicon charged particle radiation detectors. It can also be used with scintillation radiation detectors and proportional counters.
  • information +

    • Download link 671 Brochure
    • Download link 671 Brochure (A4)
    • Download link 671 Manual
  • More information +


    The Type 671 input accepts a positive or negative signal from a radiation detector preamplifier and provides a positive 0 to 10 V output signal suitable for single or multi-channel pulse height analyzers. Its gain varies continuously between 2.5 and 1500.

    Automation of key adjustments makes the Model 671 easy to set up for any radiation detector while minimizing the need for operator expertise.

    A front panel switch on the Model 671 selects triangle or Gaussian pulse shape for the UNI output connector. The noise performance of the triangular pulse shape is equivalent to the Gaussian pulse shape with a 17% longer shaping time constant. In applications where the series noise component is dominant (short shaping time constant) and a stacking rejector is used, the triangular shape usually has the same dead time and slightly lower noise than the Gaussian pulse shape. Front panel switch selectable shaping time constant for each detector and application. Six time constants range from 0.5 to 10 μs, and the TRI/GAUSS switch combination provides 12 different shaping times. Bipolar outputs are also provided for measurements requiring zero-crossing timing.

    To minimize spectral distortion at medium and high count rates, the unipolar output uses a high-performance, gated, baseline restorer with multiple automation levels. An automatic positive and negative noise discriminator ensures that the baseline restorer operates only on the true baseline between pulses and is not affected by noise level changes. It does not require the operator to adjust the baseline restorer when gain, shaping time constant, or detector characteristics change. Negative overload recovery from reset pulses generated by transistor reset preamplifiers and pulsed optical feedback preamplifiers is also handled automatically. The monitoring circuit will shut down the baseline restorer and provide a reject signal to the multi-channel analyzer until the baseline has safely recovered from the overload.

    Several operating modes can be selected for the baseline restorer. For PZ adjustment, select the PZ position. This location may also be used in applications requiring the slowest baseline recovery rate. For situations where low-frequency noise interference is a problem, high rates can be selected. On detectors where PZ cancellation is not possible, the automatic baseline restorer rate provides performance at low and high count rates.

    The unipolar output includes a front-panel limit (LIM) button to help monitor the accuracy of the PZ adjustment on the oscilloscope. When pressed, this button plugs into a diode limiter in series with the unipolar output connector. This can prevent overload distortion in the oscilloscope if a more sensitive amplitude scale is used for viewing PZ adjustments.

    An efficient stacking rejector is integrated into the Type 671 Spectral Amplifier. It provides output logic pulses to associated multi-channel analyzers to suppress spectral distortion caused by pulses stacking on top of each other at high count rates. The fast amplifier in the stack rejector includes a gated baseline restorer with automatic noise discriminator. A multi-color stack-up LED on the front panel indicates the amplifier's processing efficiency. At low count rates, the LED flashes green. The LED will turn yellow at moderate count rates and red when pulse stacking loss >70%.

    If long connecting cables are used between the detector preamplifier output and the amplifier input, cable noise caused by the environment may have an impact. The Type 671 offers two solutions. For low to moderate interference frequencies, the differential input mode can be used with paired cables from the preamplifier to suppress noise. At high frequencies, a common-mode suppression transformer built into the Type 671 input reduces noise pickup. Transformers effectively eliminate interference from display raster generators in personal computers.

    All toggle switches on the front panel can be locked to prevent accidental changes to desired settings.

  • Specifications +


    Performance

    NOTE: Unless otherwise stated, performance specifications are for use on unipolar outputs2 μsGaussian shaping sumAUTO BLRmode for measurement.

    gain range

    from2.5Arrive1500Continuously adjustable. The gain isCOARSE(coarse) andFINE GAIN(Trim Gain) control.

    Unipolar pulse
    shape

    The available switches areUNIThe output terminal selects an approximate triangular pulse shape or an approximate Gaussian pulse shape.

    bipolar output
    pulse shape

    Bipolar output pulses from0.1% rise time to amplitude is of selected shaping time1.65times. The zero-crossing time of the bipolar output pulse is delayed by the selected shaping time from the amplitude of the Gaussian unipolar output.0.33times.

    integral nonlinearity
    (UNIoutput)

    from0Arrive+10 V <±0.025%。

    noise

    Gain> 100When , the equivalent input noise<5.0μV rms, gain> 1000when,<4.5μV rms。

    temperature coefficient
    (0to50°C)

    • Unipolar output <±0.005%/°Capplies to gain,<±7.5 μV/°CSuitable for DC levels.
    • Bipolar output <±0.007%/°Capplies to gain,<±30 μV/°CSuitable for DC levels.

    swim

    in50:1Dynamic range, bipolar zero-crossing swim<±3 ns。

    Overload recovery

    Unipolar and bipolar outputs use gain in2.5within a non-overload pulse width fromX1000Overload returns to rated output2Within %.

    spectrum broadening†

    At count rates up to100,000count/seconds,FWHMwidening usually<8% while the count rate is up to200,000count/seconds, expand<15%. Under the following conditions, for60Coradioactive1.33-MeVGamma rays are measured: the efficiency is10% ofORTEC GAMMA-X PLUSDetector, unipolar output1.33-MeVThe amplitude of gamma rays is8.5V。

    spectral shift†

    At count rates up to100,000count/seconds, the peak displacement is usually<±0.018% while the count rate is up to200,000count/seconds<±0.05%. Under the conditions specified by the spectrum broadening, for1.33-MeVrays for measurement.

    Differential input

    Differential nonlinearity follows from-9 VArrive+9 V <±0.012%. Enter ±10 V(DC plus signal). common mode rejection ratio> 1000。

    Pulse accumulation rejector

    • threshold Automatically set above the noise level on the fast amplifier signal. independent of slow amplifierBLRthreshold.
    • minimum detectable signal Limited by detector and preamplifier noise characteristics.
    • Pulse pair resolution Usually500 ns. Use under the following conditions60Co 1.33-MeVGamma ray measurement: Efficiency is10% germanium detector, unipolar output1.33-MeVThe amplitude of gamma rays is4V,50,000count/seconds.

    †If measurements are made with a detector that generates a large number of slow rise time pulses or with a detector of lower quality than specified, the results may not be reproducible.













































    Controls and lights

    Fine-tune gain

    Front panel with locking dial,10Coil precision potentiometers are available0.5Arrive1.5Continuously variable direct gain coefficient.

    Coarse gain adjustment

    Front panel, eight-position switch optional5、10、20、100、200、500and1000gain coefficient.

    shaping time

    Six-position switch on front panel selects pulse shaping filter network0.5、1、2、3、6and10 μsplastic surgery time.

    mode

    The two-position locking toggle switch on the front panel can beUNI(Unipolar) Output SelectionGAUSS(Gaussian) orTRI(Triangular) pulse shaping.

    inputPOS/NEG

    Front-panel, two-position locking toggle switch selects positive or negative input polarity.

    NORM/DIFF

    Two-position slide switch mounted on printed circuit board with optional standard (NORM) or differential (DIFF) input mode. inNORMLocation, front and rear panelsINPUTThe connector serves as a standard input for two identical preamplifier signal cables. inDIFFmode, rear panelINPUTconnector becomes the differential ground reference input, and the front panelINPUTStill the standard input for preamp signal cables. inDIFFmode, the preamp signal cable is connected to the front panelINPUT, while another cable whose center conductor is connected to the preamplifier ground via an impedance matching resistor is connected to the rear panelINPUT. The impedance matching resistor must match the output impedance of the preamplifier.

    BAL(differential input gain balanced)

    installed inside the modulePCon the board20The loop potentiometer matches the gain of the standard and differential reference inputs to achieveDIFFCommon mode noise rejection in mode.

    PZadjust

    on the front panel20The loop potentiometer can be adjusted with a screwdriverPZCancel. Adjust coverage40 μsPreamplifier exponential decay time constant to ∞. For transistor reset preamplifiers or pulsed optical feedback preamplifiers, set counterclockwisePZAdjust.

    LIMbutton

    Plug in and front panelUNIdiode limiter in series with the output connector. Observe within a more sensitive oscilloscope rangePZAdjusting the accuracy prevents overload distortion in the oscilloscope.

    BLR

    Front panel, three-position locking, toggle switch can be used to select the baseline recovery rate.PZThe fixed rate provided by the position, used to adjustPZCancel. Automatic position at low count rates withPZThe position's rate matches, but as the count rate increases, the recovery rate will also increase. The high-rate position is used to suppress low-frequency interference.

    PURaccept/rejectLED

    multicolorLEDIndicates the percentage of pulses that were rejected due to pulse pile-up. greenLEDexpress0-40%, yellow indicates40-70%, red indicates rejection> 70%。


    Input

    Input (front panel)

    BNCThe connector accepts preamplifier signals of either polarity with a rise time less than the selected shaping time and an exponential decay time constant of40 μsto ∞. forNEGEnter the switch setting and the coarse gain is5When the input impedance is1000 Ω, coarse gain setting ≥10When the input impedance is465 Ω. forPOSEnter the switch setting and the coarse gain is5When the input impedance is2000 Ω, coarse gain ≥10When the input impedance is1460 Ω. The inputs are DC coupled and protected to ±25 V。

    Input (rear panel)

    BNCconnector. WhenPWBinstalledNORM/DIFFThe slide switch is inNORMposition, with the front panelINPUTSame. When the slide switch is set toDIFF, in differential input mode, the rear panelINPUTGround reference connector for preamplifier. forDIFFandPOSEnter the switch setting and the coarse gain is5When the input impedance is1000 Ω, coarse gain setting ≥10When the input impedance is465 Ω. forDIFFandNEGEnter the switch setting and the coarse gain is5When the input impedance is2000 Ω, coarse gain ≥10When the input impedance is1460 Ω. Inputs are DC coupled; protected to ±25 V。

    INH IN

    rear panelBNCThe input connector is inhibited from accepting a reset signal from a transistor reset preamplifier or a pulsed optical feedback preamplifier. You can use the correctNIMStandard logic pulse orTTLlevel. Logic is selectable as active high or active low via PC board jumpers. Disabling the input will activate protection against distortion caused by preamplifier reset. This includes turning off the baseline restorer, monitoring the unipolar output for negative overload recovery, and generatingPUR(rejection) andBUSY(Busy) signal.PURandBUSY(Busy) logic pulse is used in the relevantADCOr prevent analysis and correction reset dead time in multi-channel analyzers.














    Output

    UNI

    Front and rear panelsBNCThe connector provides positive, unipolar pulses with a linear output range of0to+10 V. Front panel output impedance<1Ω. Using PCB jumpers, optional<1Ω or93Rear panel output impedance in Ω. The output DC returns to0±5 mVAnd has short circuit protection function.

    BI

    Front and rear panelsBNCThe connector provides bipolar pulses with predominantly positive lobes. The linear output range is0to±10 V. Front panel output impedance<1Ω. Using PCB jumpers, optional<1Ω or93Rear panel output impedance in Ω. The baseline between pulses has0±10 mVDC level. Short circuit protection.

    CRM

    The rate meter output has a rear panelBNCConnector is provided for each linear input pulse that exceeds the stackup detector threshold250 nswide+5 Vlogic signal. The output impedance is50Ω。

    busy

    rear panelBNCThe connector remains active for the duration that the linear signal exceeds the positive or negative baseline restorer threshold or stacking detector threshold or duringINH INThe input signal is provided for the duration of the+5 Vlogic pulse. This applies to relatedADCor dead time correction for multi-channel analyzers. RightNIMStandard logic pulses are selectable as active high or active low via printed circuit board jumpers. The output impedance is50Ω。

    PUR

    The stack rejection output is a rear panelBNCconnector. When pulse build-up is detected, it will provide+5 V NIMStandard logic pulse. A pulse reset preamplifier's output exists during reset and resets overload recovery. The output pulses are selectable as active high or active low via printed circuit board jumpers. The output impedance is50Ω. related toADCor used in conjunction with a multi-channel analyzer to prevent analysis of distorted pulses.

    preamplifier

    rear panel standardORTECconnector(Amphenol 17-10090) to power the associated preamplifier. with all standardsORTECPower cord compatible on preamplifier.


























    electrical and mechanical

    Required power supply

    671Model consists of ±24 Vand±12 VofNIMpower box (e.g.ORTEC 4001A/4002Atype power box/power supply). The required power supply is+24 V,100 mA;-24 V,200 mA;+12 V,325 mAand-12 V,180 mA。

    weight

    Net weight 1.5Kilogram (3.3pounds). Shipping weight 3.1Kilogram (7.0pounds).

    Size

    Standard single widthNIMmodule,3.43 x 22.13cm (1.35 x 8.714inches) front panel, conforming toDOE/ER-0457Trequirements.